Technical specifications - Expert Transient
All data overview
Analog inputs | 4, 8, 12 or 16 |
Sampling rate, adjustable per channel | 20 Hz .. 50 kHz |
Usable signal range | DC .. 20 kHz |
Voltage range | ± 25 V |
Current measuring range | 0 .. 20 mA / 4 .. 20 mA / free |
Signal conditioning, software switchable | None, AC coupling, IEPE |
Resolution/type ADC / input impedance | 24 bit Sigma-Delta / 4 GΩ |
Dielectric withstand voltage | ± 100 VDC |
Galvanic isolation channel to channel / channel to system / PE | ± 400 VDC |
Measurement accuracy | 0.5 mV + 0.008 % from measurement value |
Digital inputs | 4 |
Input signal | low: 0 .. 1 V / high: 5 .. 100 VDC@3.5 mA |
Galvanic isolation channel to channel / channel to system / PE | ± 400 VDC |
Max. input frequency / min. pulse width | 1 MHz / 500 ns |
Analog outputs | 4 |
Resolution | 16 bit |
Output ranges | 0 .. 10 V / ± 10 V / 0 .. 20 mA / 4 .. 20 mA / ±20 mA |
Galvanic isolation channel to channel / channel to system / PE | ± 400 VDC |
Min. load / max. burden | 650 Ω |
Digital outputs | 8 |
Max. switching voltage / current / PWM | 50 V / 2.5 A for 2 seconds, 250 mA continuous current / galvanically isolated / 5 Hz .. 10 kHz, up to 1:1000 |
Data storage | |
Internal / external | 32 GB (approx. 2,300 million measured values) / USB, NFS, CIFS, (S)FTP |
Signal comditioning functions | |
High-pass / low-pass / bandpass filter | |
Cut-off frequency / filter order / filter characteristic | 5 Hz .. 20 kHz / 4, 6, 8, 10 / incl. Butterworth |
Analysis | |
Integrator / Differentiator | Single or double integrator / differentiator |
Characteristic values from time signal | |
Maximum / minimum value, peak-to-peak value, arithm. average, TRMS, max, vect. sum, local min. / max. values | |
Characteristic values on digital input | |
Frequency | 0.2 Hz .. 1 MHz |
Counter | 64 bit, UP-Down counter, quadratur decoder |
Interfaces | |
Physical equipment COM 1 / COM 2 | RS485, 9-pole sub-D socket, IEC 61158-2 |
Physical equipment COM 3 | RS232, 9-pole sub-D-plugs |
LAN | 1 x 1000BASE-T |
USB | Device 2.0 / Host 2.0 / low / high / full |
PROFIBUS | 2 x PROFIBUS DPV1 / Slave max. 12 Mbit, also redundant according to PNO 2.212 V1.2, passive PROFIBUS sniffer |
CAN / RS 232/485 | 2 x CAN 2.0 / Modbus RTU Master & Slave, SCPI, ASCII |
TCP/IP | Modbus TCP Server & Client, OPC UA Server & Client, OPC UA HA Server (optional), SMTP, NTP, PTP |
General technical information | |
Dimensions / weight | W 217 mm x H 88 mm x D 128 mm / 750 g |
Fixing | Mounting rail DIN EN 60715 or screw mounting |
Connecting terminals | Plug-in screw terminals, 96 terminals in 2 rows, max. 1.5 mm² |
Temperature range | -20 .. 50 °C* |
Humidity | max. 90 % relative humidity, non-condensing |
Supply voltage | 12 .. 24 V / ± 10% |
Power input | max. 20 W |
* Sufficient air circulation in the control cabinet must be ensured.
Status: 27.09.2024